20130424000000162572.jpg

발광측정기

상담·문의하기

Microplate Reader System

  • 모델명 Synergy H1
  • 제조사 Biotek
  • 장비용도 교육
  • 장비구분 화합물전처리/분석장비 > 바이오제조/분석장비 > 마이크로플레이트리더

상세정보

구축일자 2012-08-23
납품업체 Biotek
위치 광주광역시 북구 첨단과기로 123 (오룡동) 지스트대학 지스트대학 B동 F4 412호
구성및성능 General
Detection mode Fluorescence Luminescence UV-Visible Absorbance
Read method End point kinetic spectral scanning well-area scanning
Microplate types 6- to 1536-well plates (luminescence 1-384)
PCR plates
Compatible with Take3 Multi-Volume Plate with 2 μL microspots
Temperature control 4°C above ambient to 65°C; +0.5°C at 37°C
Shaking Yes
Software Gen5™ included
Control through USB or serial port
Automation Yes
Light source Tungsten Halogen Lamp
Sensitivity 100 amol of biotinylated-LCK-P peptide 25 μL/well in 384-well plate
Dynamic range >6 decades
Reading speed 2 minutes typical for 96-well plate
Detection system Ultra Low Noise PMT
Absorbance
Light source SQ Xenon Flash
Wavelength selection Monochromator
Wavelength range 200 - 999 nm 1 nm increment
Bandpass 2.4 nm
Dynamic range 0 - 4.0 OD
Resolution 0.0001 OD
Pathlength correction Yes
Monochromator wavelength accuracy +2 nm
Monochromator wavelength repeatability +0.2 nm
OD accuracy <1% at 2 OD typical
<3% at 3.0 OD typical
OD li
장비안내 The Synergy H1 Hybrid Reader is BioTek's high-performance multi-mode plate reader designed for the life science research and drug discovery markets. Its fluorescence detection system uses deep blocking filters and dichroic mirrors for the best level of performance in research and screening applications. The dedicated absorbance detection system is monochromator-based providing the best level of flexibility in this read mode. And by utilizing dedicated optical elements for each individual detection technique Synergy 2 offers uncompromised performance and the lowest limits of detection.

Detection modes include Fluorescence Intensity Luminescence and UV-visible Absorbance. Its three broad-spectrum light sources have been chosen for optimal illumination and excitation in all applications.
Measurement Techniques: Fluorescense Intensity Glow and Flash Luminescence UV-Vis Absorbance FRET TR-FRET BRET Area Scanning Spectral Scanning
Unique Optical System: a narrow band monochromator optimized filter/dichroic mirror