Abstract
STATEMENT OF PROBLEM. A few of studies which compared and continuously measured the stability of various surface treated implants
in the same individual had been performed. PURPOSE. We aim to find the clinical significance of surface treatments by observing the differences in the stabilization stages of implant stability. MATERIAL AND METHODS. Eight different surface topographies of dental implants were especially designed for the present study. Machined surface implants were used as a control group. 4 nano-treated surface implants (20 nm TiO2 coating surface, heat-treated 80 nm TiO2 coating surface, CaP coating surface, heat treated CaP coating surface) and 3 micro-treated surface implants [resorbable blast media (RBM) surface, sandblast and acid-etched (SAE) surface, anodized RBM surface] were used as experiment groups. All 24 implants were placed in 3 adult dogs. Periotest? & ISQ values measured for 8 weeks and all animals were sacrificed at 8 weeks after surgery. Then the histolog