토양 내 구리분석에 LIBS와 ICP 비교
Comparisons between the ICP and LIBS Analyses of Copper in Soils
한국환경분석학회
박정훈
논문정보
- Publisher
- 한국환경분석학회지
- Issue Date
- 2015-12-01
- Keywords
- -
- Citation
- -
- Source
- -
- Journal Title
- -
- Volume
- 18
- Number
- 4
- Start Page
- 212
- End Page
- 220
- DOI
- ISSN
- 12269700
Abstract
Laser Induced Breakdown Spectroscopy (LIBS) has been introduced as a type of analytical equipment to determine elements on the surface of various materials. In this study, standard and sample soils were molded to make a pellet for LIBS analysis. The soils were also digested with acids for ICP analysis. Accuracies ranging between 86.5% and 119.9%, an RSD of 12.3%, an LOD of 6.09 mg/kg, and an LOQ 19.41 mg/kg resulted from the analysis of copper in soils using LIBS. Correlations (R2) of LIBS with ICP-aqua regia digestion and with ICP-total digestion analysis concentrations were over 0.96 for 28 samples. The total analytical time for the LIBS method was 385 min, which was 43.4% shorter than for the ICP-OES method (680 min). For the analysis of copper in soils, LIBS can be used for on-site detection without chemical preparation.
- 전남대학교
- KCI
- 한국환경분석학회지
저자 정보
| 이름 | 소속 |
|---|---|
| 박정훈 | 환경에너지공학과 |