Research Hub

대학 자원

대학 인프라와 자원을 공유해 공동 연구와 기술 활용을 지원합니다.

Loading...

논문 리스트

2017
광학식측정장치를 이용한 금속정 내고정 수술의 원위부 나사체결을 위한 로보틱 유도 시스템 Robotic Guidance of Distal Screwing for Intramedullary Nailing Using Optical Tracking System
한국로봇학회
고성영
논문정보
Publisher
로봇학회 논문지
Issue Date
2017-12-01
Keywords
-
Citation
-
Source
-
Journal Title
-
Volume
12
Number
4
Start Page
411
End Page
418
DOI
ISSN
19756291
Abstract
During the intramedullary nailing procedure, surgeons feel difficulty in manipulation of the X-ray device to align it to axes of nailing holes and suffer from the large radiation exposure from the X-ray device. These problems are caused by the fact the surgeon cannot see the hole’s location directly and should use the X-ray device to find the hole’s location and direction. In this paper, we proposed the robotic guidance of the distal screwing using an optical tracking system. To track the location of the hole for the distal screwing, the reference marker is attached to the proximal end of an intramedullary nail. To guide the drill’s direction robustly, the 6-degree-of-freedom robotic arm is used. The robotic arm is controlled so as to align the drill guiding tool attached the robotic arm with the obtained the hole’s location. For the safety, the robot’s linear and angular velocities are restricted to the predefined values. The experimental results using the artificial bones showed that the position error and the orientation error were 0.91 mm and 1.64°, respectively. The proposed method is simple and easy to implement, thus it is expected to be adopted easily while reducing the radiation exposure significantly.

저자 정보

이름 소속
고성영 기계공학부